TY - THES AB - This experiment investigates the effects of using electron beam Cerrobend contact skin collimators in combination with thin metallic scatter foils on electron beam depth dose distribution and isodose curves for small electron fields. The premise is to provide a clinically more uniform treatment for small superficial lesions. This is a test of concept that could potentially open the door for the development of novel electron beam therapy technique applicable for clinical use. The ultimate goal is to produce a quickly accessed, easy to fabricate, reusable device for clinical use with current linear accelerators without the need to modify the design of current linear accelerators? built-in scatter foils or make any other changes to the machine. AU - Alhussan, Abdulaziz S.A. DA - 2020 DO - 10.6083/6682x449n DO - DOI ID - 8575 KW - Skin Neoplasms KW - skin collimation KW - isodose curves KW - electron therapy L1 - https://digitalcollections.ohsu.edu/record/8575/files/Alhussan.Abdulaziz.2020.pdf L2 - https://digitalcollections.ohsu.edu/record/8575/files/Alhussan.Abdulaziz.2020.pdf L4 - https://digitalcollections.ohsu.edu/record/8575/files/Alhussan.Abdulaziz.2020.pdf LK - https://digitalcollections.ohsu.edu/record/8575/files/Alhussan.Abdulaziz.2020.pdf N2 - This experiment investigates the effects of using electron beam Cerrobend contact skin collimators in combination with thin metallic scatter foils on electron beam depth dose distribution and isodose curves for small electron fields. The premise is to provide a clinically more uniform treatment for small superficial lesions. This is a test of concept that could potentially open the door for the development of novel electron beam therapy technique applicable for clinical use. The ultimate goal is to produce a quickly accessed, easy to fabricate, reusable device for clinical use with current linear accelerators without the need to modify the design of current linear accelerators? built-in scatter foils or make any other changes to the machine. PB - Oregon Health and Science University PY - 2020 T1 - Exploring the effects of novel design of contact lead sheet collimation and filtration of high energy electron fields on electron depth dose distributions with application to uniform treatment of small skin lesions TI - Exploring the effects of novel design of contact lead sheet collimation and filtration of high energy electron fields on electron depth dose distributions with application to uniform treatment of small skin lesions UR - https://digitalcollections.ohsu.edu/record/8575/files/Alhussan.Abdulaziz.2020.pdf Y1 - 2020 ER -