TY - THES N2 - The lungs are the primary dose limiting organ in high dose total body irradiation (TBI). Accurate prediction of lung dose can dictate treatment success by preventing pulmonary toxicity. Normal treatment planning consists of basic hand calculations and potentially in vivo dosimetry. The use of a treatment planning system (TPS) is an unverified method at the extended treatment distance required by TBI. In this study, we attempted to analyze the Eclipse TPS for used in a TBI setup. Both its ability to determine the required monitor units (MU) and its ability to compensate for tissue heterogeneity in lung dose estimates were compared to traditional techniques and dosimetry. DO - 10.6083/2j62s5533 DO - DOI AB - The lungs are the primary dose limiting organ in high dose total body irradiation (TBI). Accurate prediction of lung dose can dictate treatment success by preventing pulmonary toxicity. Normal treatment planning consists of basic hand calculations and potentially in vivo dosimetry. The use of a treatment planning system (TPS) is an unverified method at the extended treatment distance required by TBI. In this study, we attempted to analyze the Eclipse TPS for used in a TBI setup. Both its ability to determine the required monitor units (MU) and its ability to compensate for tissue heterogeneity in lung dose estimates were compared to traditional techniques and dosimetry. T1 - Validation of treatment planning and lung dose estimation for total body irradiation using the eclipse treatment planning system DA - 2022 AU - Heath, Mason W. L1 - https://digitalcollections.ohsu.edu/record/9738/files/Heath.Mason.2022.pdf PB - Oregon Health and Science University PY - 2022 ID - 9738 L4 - https://digitalcollections.ohsu.edu/record/9738/files/Heath.Mason.2022.pdf KW - Planning Techniques KW - Whole-Body Irradiation KW - In Vivo Dosimetry KW - Motivation KW - Radiotherapy TI - Validation of treatment planning and lung dose estimation for total body irradiation using the eclipse treatment planning system Y1 - 2022 L2 - https://digitalcollections.ohsu.edu/record/9738/files/Heath.Mason.2022.pdf LK - https://digitalcollections.ohsu.edu/record/9738/files/Heath.Mason.2022.pdf UR - https://digitalcollections.ohsu.edu/record/9738/files/Heath.Mason.2022.pdf ER -