Development of a standard linear accelerator based back-up plan for total marrow irradiation with helical tomography Public Deposited

Purpose: To identify a field match line for TMI treatment interruption, and to develop a total body irradiation (TBI) completion plan to be used in the case of a treatment interruption during a total marrow irradiation (TMI) treatment with helical tomotherapy. Conclusion: Finding a match line from a TMI treatment interruption using film dosimetry was determined to be feasible. A completion procedure using a standard TBI setup is feasible, although further investigation into field size feathering is required in order to achieve adequate dose homogeneity in the junction region.


Creator
Publisher
Identifier
  • morelli.samantha.2019.pdf
  • https://doi.org/10.6083/m326m224z
Keyword
Publication Date
  • 2019
Document type

Items